Repositori DSpace/Manakin

A correction procedure for the asymmetry of differential pressure transducers in respiratory impedance measurements

Mostra el registre parcial de l'element

dc.creator Farré Ventura, Ramon
dc.creator Navajas Navarro, Daniel
dc.creator Peslin, René
dc.creator Rotger i Estapé, Maria del Mar
dc.creator Duvivier, Claude
dc.date 2009-06-17T08:58:18Z
dc.date 2009-06-17T08:58:18Z
dc.date 1989
dc.date.accessioned 2024-12-16T10:18:23Z
dc.date.available 2024-12-16T10:18:23Z
dc.identifier 0018-9294
dc.identifier http://hdl.handle.net/2445/8687
dc.identifier 35255
dc.identifier.uri http://fima-docencia.ub.edu:8080/xmlui/handle/123456789/7610
dc.description The usual setup for measuring respiratory input impedance requires a differential pressure transducer attached to a pneumotachograph. Because no data correction procedure has been devised to account for transducer asymmetry, a highly symmetrical transducer is required to obtain reliable impedance data. Here, a general model for the measuring system is presented. Its main feature is that differential pressure transducers are modeled as two-input-one-output systems. From the theoretical model, a dynamic calibration and data correction procedure is defined. This was tested using highly asymmetrical transducers (common-mode rejection ratio between 45 and 27 dB) to measure the impedance of two respiratory analogs. Results obtained show that respiratory input impedance can be adequately measured if data are corrected for transducer asymmetry.
dc.format 4 p.
dc.format application/pdf
dc.language eng
dc.publisher IEEE
dc.relation Reproducció del document publicat a http://dx.doi.org/10.1109/10.40822
dc.relation IEEE Transactions on Biomedical Engineering, 1989, vol. 36, núm. 11, p. 1137-1140.
dc.relation http://dx.doi.org/10.1109/10.40822
dc.rights (c) IEEE, 1989
dc.rights info:eu-repo/semantics/openAccess
dc.source Articles publicats en revistes (Ciències Fisiològiques)
dc.subject Mesurament
dc.subject Errors científics
dc.subject Biomedical measurement
dc.subject Measurement errors
dc.subject Modeling
dc.subject Pneumodynamics
dc.title A correction procedure for the asymmetry of differential pressure transducers in respiratory impedance measurements
dc.type info:eu-repo/semantics/article
dc.type info:eu-repo/semantics/publishedVersion


Fitxers en aquest element

Fitxers Grandària Format Visualització

No hi ha fitxers associats a aquest element.

Aquest element apareix en la col·lecció o col·leccions següent(s)

Mostra el registre parcial de l'element

Cerca a DSpace


Cerca avançada

Visualitza

El meu compte